CA LV124, Calibration Device for OEM LV124 and OEM LV148

Brand:

EM TEST

CA LV124, Calibration Device for OEM LV124 and OEM LV148
  • Product Description
  • Main features
  • Technical parameters
    • Commodity name: CA LV124, Calibration Device for OEM LV124 and OEM LV148
    • Commodity ID: 1064615360851496960
    • 品牌11: EMTEST
    • 行业: 汽车电子和整车
    • 描述: The LV 124 defines a reference measurement, each of which is 100 ohms (± 5%) and 1 ohm (± 5%), which must be performed and recorded as a DUT replacement.
    • 品牌: EM TEST

    LV 124 defines a reference measurement, each of 100 ohms (±5%) and 1 ohm (±5%), that must be performed and documented as a DUT replacement. This test setup must provide verification of edge steepness. The low inductance part must be used as a resistor.

  • Standards compliant:

      BMWGS 95024-2-1

      MBN LV 124-1

      OEM LV 124

      OEM LV 148

      VW 80000

    Power cord test steps:

      LV 124 defines a reference measurement, each of 100 ohms (±5%) and 1 ohm (±5%), that must be performed and documented as a DUT replacement. This test setup must provide verification of edge steepness. The low inductance part must be used as a resistor.

      LV 148 defines equivalent reference measurements for 1000 ohms and 10 ohms.

      The load resistor must have the following characteristics to achieve the desired pulse parameters:

      - Low inductance components

      - Shortest cable to PFM output (a few centimeters)

      - The probe must be placed as close as possible to the load resistor, or as with the CA LV 124-P, directly at the load using a BNC connection.

      

    EMTEST

     

    Data line test steps:

      A reference measurement must be made and documented using 1 kohm (±5%) and 1 ohm (±5%) as surrogates for the DUT. This test setup must provide verification of edge steepness. The low inductance part must be used as a resistor.

      During the test, all even and odd I/O lines are alternately toggled simultaneously. Ensure that only one line is being charged at a time by testing the adapter.

      

    EMTEST

     

    controlling software:

      Power cord settings

      Selecting Edge Verification in the software allows selection of reference measurements. There is a test window for recalling the reference measurement setup.

      1: Edge Verification: Select this option to perform a reference measurement.

      2: Test window for reference measurement

      

    EMTEST

     

      Data line settings

      Selecting Edge Verification in the software allows selection of reference measurements. There is a test window for recalling the reference measurement setup.

      1: Edge Verification: Select this option to perform a reference measurement.

      2: Test window for reference measurement

      

    EMTEST

  • Specific indicators:

    CA LV 124 Components

    CA LV124-P1R

    Power Line Load Resistance

    CA LV124-P100R

    Power Line Load Resistance

    CA LV124-D1R

    Data Line Load Resistance

    CA LV124-D1000R

    Data Line Load Resistance

    BNC cable

    BNC measurement cable 2m, power line verification

    power cable

    Red/black 2 wires, 0.5m long, signal data wire verification

    CA LV 148 Components

    CA LV148-P10R

    Power Line Load Resistance

    CA LV148-P1000R

    Power Line Load Resistance

    BNC cable

    BNC measurement cable 2m, power line verification

    CA LV 124-P1R

    Wire

    power cable

    Resistance

    1 ohm

    Accuracy

    ± 1 %

    maximum voltage

    12 V

    Maximum power

    30 W, (150 W, 1 s peak)

    size

    185 mm x 105 mm x 58 mm

    weight

    1.05 kg

    CA LV 124-P100R

    Wire

    power cable

    Resistance

    100 ohm

    Accuracy

    ± 1 %

    maximum voltage

    100 V

    Maximum power

    30 W, (100 W, 1 s peak)

    size

    185 mm x 105 mm x 58 mm

    weight

    0.90 kg

    CA LV 124-D1R

    Wire

    data line

    Resistance

    1 ohm

    Accuracy

    ± 2%

    maximum voltage

    2 V

    Maximum power

    1 W, (4W, 1 s peak)

    size

    120 mm x 65 mm x 40 mm

    weight

    0.15 kg

    CA LV 124-D1000R

    Wire

    data line

    Resistance

    1000 ohm

    Accuracy

    ± 2%

    maximum voltage

    40 V

    Maximum power

    1 W, (4W, 1 s peak)

    size

    120 mm x 65 mm x 40 mm

    weight

    0.15 kg

    general data

    temperature

    10 °C to 35 °C  

    Relative humidity

    25 % - 75 %, non-condensing

    atmospheric pressure

    86 kPa  - 106 kPa