EMC Test System For Civil Products
-
- Electrostatic Discharge Immunity
- Radiated, radio-frequency,electromagnetic field immunity
- Electrical Fast Transient Burst Immunity
- Surge immunity
- Immunity To Conducted Disturbance Induced by Radio Frequency Field
- Power Frequency Magnetic Field Immunity
- Voltage dips, short interruptions and voltage variations immunity
- Harmonics and interharmonics including mains signalling at AC power port, low frequency immunity
- Voltage Fluctuation Immunity Test
- Common mode disturbances in the frequency range 0 Hz to 150 kHz Immunity
- Ripple on DC input power port immunity
- Three-phase Voltage Unbalance Immunity Test
- Power Frequency Variation Immunity Test
- Oscillatory Wave Immunity Test
- Damped Oscillatory Magnetic Field Immunity Test
- Differential mode disturbances immunity test
- DC power input port voltage dip, short interruption and voltage variations test
-
Automotive Electronic EMC Test System
-
- Electrostatic Discharge Immunity
- Electrical Transient Conducted Immunity
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Anechoic Chamber Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Transverse Wave (TEM) Cell Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-large Current injection (BCI) method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Stripline Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-direct Injection Of Radio Frequency (RF) Power
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Magnetic Field Immunity Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Portable Transmitter Simulation Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Conduction Immunity Method For Extended Audio Range
- High Voltage Electrical Performance ISO 21498-2 Test System
- High Voltage Transient Conducted Immunity (ISO 7637-4)
-
-
- CE101(25Hz ~ 10kHz power line conduction emission)
- CE102(10kHz ~ 10MHz power line conduction emission)
- CE106(10kHz ~ 40GHz antenna port conducted emission)
- CE107 (Power Line Spike (Time Domain) Conducted Emission)
- RE101(25Hz ~ 100kHz magnetic field radiation emission)
- RE102(10kHz ~ 18GHz electric field radiation emission)
- RE103(10kHz ~ 40GHz antenna harmonic and spurious output radiated emission)
-
- CS101(25Hz ~ 150kHz power line conduction sensitivity)
- CS102(25Hz ~ 50kHz ground wire conduction sensitivity)
- CS103(15kHz ~ 10GHz Antenna Port Intermodulation Conducted Sensitivity)
- CS104(25Hz ~ 20GHz antenna port unwanted signal suppression conduction sensitivity)
- CS105(25Hz ~ 20GHz antenna port intermodulation conduction sensitivity)
- CS106 (Power Line Spike Signal Conduction Sensitivity)
- CS109(50Hz ~ 100kHz shell current conduction sensitivity)
- CS112 (Electrostatic Discharge Sensitivity)
- CS114(4kHz ~ 400MHz cable bundle injection conduction sensitivity)
- CS115 (Conduction sensitivity of cable bundle injection pulse excitation)
- CS116(10kHz to 100MHz Cable and Power Line Damped Sinusoidal Transient Conduction Sensitivity)
- RS101(25Hz ~ 100kHz magnetic field radiation sensitivity)
- RS103(10kHz ~ 40GHz electric field radiation sensitivity)
- RS105 (Transient Electromagnetic Field Radiated Susceptibility)
-
-
-
-
-
-
-
-
-
-
EMC Test System For Civil Products
-
- Electrostatic Discharge Immunity
- Radiated, radio-frequency,electromagnetic field immunity
- Electrical Fast Transient Burst Immunity
- Surge immunity
- Immunity To Conducted Disturbance Induced by Radio Frequency Field
- Power Frequency Magnetic Field Immunity
- Voltage dips, short interruptions and voltage variations immunity
- Harmonics and interharmonics including mains signalling at AC power port, low frequency immunity
- Voltage Fluctuation Immunity Test
- Common mode disturbances in the frequency range 0 Hz to 150 kHz Immunity
- Ripple on DC input power port immunity
- Three-phase Voltage Unbalance Immunity Test
- Power Frequency Variation Immunity Test
- Oscillatory Wave Immunity Test
- Damped Oscillatory Magnetic Field Immunity Test
- Differential mode disturbances immunity test
- DC power input port voltage dip, short interruption and voltage variations test
-
Automotive Electronic EMC Test System
-
- Electrostatic Discharge Immunity
- Electrical Transient Conducted Immunity
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Anechoic Chamber Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Transverse Wave (TEM) Cell Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-large Current injection (BCI) method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Stripline Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-direct Injection Of Radio Frequency (RF) Power
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Magnetic Field Immunity Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Portable Transmitter Simulation Method
- Immunity Test To Narrowband Radiated Electromagnetic Energy-Conduction Immunity Method For Extended Audio Range
- High Voltage Electrical Performance ISO 21498-2 Test System
- High Voltage Transient Conducted Immunity (ISO 7637-4)
-
-
- CE101(25Hz ~ 10kHz power line conduction emission)
- CE102(10kHz ~ 10MHz power line conduction emission)
- CE106(10kHz ~ 40GHz antenna port conducted emission)
- CE107 (Power Line Spike (Time Domain) Conducted Emission)
- RE101(25Hz ~ 100kHz magnetic field radiation emission)
- RE102(10kHz ~ 18GHz electric field radiation emission)
- RE103(10kHz ~ 40GHz antenna harmonic and spurious output radiated emission)
-
- CS101(25Hz ~ 150kHz power line conduction sensitivity)
- CS102(25Hz ~ 50kHz ground wire conduction sensitivity)
- CS103(15kHz ~ 10GHz Antenna Port Intermodulation Conducted Sensitivity)
- CS104(25Hz ~ 20GHz antenna port unwanted signal suppression conduction sensitivity)
- CS105(25Hz ~ 20GHz antenna port intermodulation conduction sensitivity)
- CS106 (Power Line Spike Signal Conduction Sensitivity)
- CS109(50Hz ~ 100kHz shell current conduction sensitivity)
- CS112 (Electrostatic Discharge Sensitivity)
- CS114(4kHz ~ 400MHz cable bundle injection conduction sensitivity)
- CS115 (Conduction sensitivity of cable bundle injection pulse excitation)
- CS116(10kHz to 100MHz Cable and Power Line Damped Sinusoidal Transient Conduction Sensitivity)
- RS101(25Hz ~ 100kHz magnetic field radiation sensitivity)
- RS103(10kHz ~ 40GHz electric field radiation sensitivity)
- RS105 (Transient Electromagnetic Field Radiated Susceptibility)
-
-
-
-
-
-
Product
EMC Testing Solution Provider
ESR EMI measurement receiver
Brand:
ROHDE&SCHWARZ
- Product Description
- Main features
- Technical parameters
-
- Commodity name: ESR EMI measurement receiver
- Commodity ID: 1064827469841453056
- 品牌11: ROHDE&SCHWARZ
- 描述: R&S ® ESR is an EMI testing receiver with a frequency range of 10 Hz to 26.5 GHz and complies with the CISPR 16-1-1 standard.
- 品牌: ROHDE&SCHWARZ
The R&S®ESR is an EMI test receiver with a frequency range from 10 Hz to 26.5 GHz and complies with CISPR 16-1-1. The receiver can measure electromagnetic interference with conventional stepped frequency sweeps or at very high speeds with FFT-based time domain sweeps. At the same time, the R&S®ESR can also be used as a comprehensive and powerful signal and spectrum analyzer for laboratory applications. With real-time spectrum analysis capabilities and extensive diagnostic tools, the test receiver can be used to analyze interfering signals and their history in detail. The R&S®ESR features clearly structured menus and an intuitive touchscreen for easy operation in any mode.
-
EMI test receiver and signal/spectrum analyzer in one
Complies with CISPR 16-1-1 version 3.1
Preselection with integrated 20 dB preamplifier
The resolution bandwidth is CISPR compliant, with the option of 10 Hz to 1 MHz
Decimal steps (MIL STD-461, DO-160)
Ultra-fast time domain sweep (optional) or traditional stepped-frequency sweep
Real-time spectrum analysis across up to 40 MHz for detailed investigation of interference (optional)
High-resolution time domain display (50 μs)
Automated test programs
IF analysis (optional)
Features and benefits:
Standard-compliant EMI measurements
Certified measurements.
Perform standard-compliant EMI measurements in spectrum analyzer mode.
Ultra-fast measurements with FFT-based time domain scanning (option)
Because of the hardware-based FFT calculation, the measurement speed is extremely fast
Conducted interference levels are measured in real time by synchronous quasi-peak and average-weighted
Real-time spectrum analysis for detailed investigation of interference (option)
Seamlessly display the spectrum in the time domain with a waterfall chart
Afterglow mode (spectral histogram) is used to clearly distinguish between pulsed and continuous interference
Frequency templates are triggered to accurately and reliably detect irregular spectral events
Powerful in-frequency and intra-time measurement and analysis capabilities
Automated test sequences covering preview measurements, data simplification, and final measurements
IF analysis function to display the spectrum around the interference signal (optional)
Time Domain Display – Integrated oscilloscope functionality
Synchronously displays up to six traces along with four bar charts
Preset antenna factors (conversion factors) and user-created sets of conversion factors
Business-standard EMI limit line library with quick editor
Remote control of measurements and automated EMI test procedures using R&S EMC32 measurement software and R&S®®ES-SCAN EMI software
Extensive analytical capabilities for general laboratory applications
Trace generator for scalar network analysis (optional)
Easy to operate, display information is easy to understand
Touchscreen-based user interface with undo/redo functionality
Visualize scan sheets
Integrate online help
Save results and instrument settings to internal or external media
Removable hard disk drivers (HDDs) to protect the confidentiality of test data
Remote control via GPIB or LAN
Drivers for LabView, LabWindows/CVI, VXI Plug & Play
Firmware updates – Keep up with new developments
Compact and robust – also suitable for mobile applications
DC power supply for field applications with optional external battery compartment and ruggedized housing
Solid-state drives (SSDs) resistant to high vibration and shock loads
Compact design
-
frequency
Frequency Range
R&S®ESR3
9 kHz to 3.6 GHz
R&S®ESR3 ( R&S®ESR-B29 )
10 Hz to 3.6 GHz
R&S®ESR7
9 kHz to 7 GHz
R&S®ESR7 ( R&S®ESR-B29 )
10 Hz to 7 GHz
R&S®ESR26
9 kHz to 26.5 GHz
R&S®ESR26 R&S®ESR-B29 )
10 Hz to 26.5 GHz
Level
Maximum RF level (CW)
RF attenuation ≥ 10 dB;
Turn off the RF preamplifier
RF attenuation ≥ 10 dB;
Turn on the RF preamplifier30 dBm (= 1 W)
23 dBm (= 0.2 W)Maximum pulse voltage
RF attenuation ≥ 10 dB
Enter 1
Enter 2
150 V
450 VMaximum pulse energy
RF attenuation ≥ 10 dB; 10 μs
Enter 1
Enter 2
1 mWs
20 mWs1 dB compress
RF attenuation 0 dB; Turn off the RF preamplifier and preselection
+3 dBm, the nominal value
IF and resolution bandwidth
Analyzer mode (frequency span ≥ 10 Hz) and
Receiver mode10 Hz to 10 MHz (–3 dB) in 1/2/3/5/10 steps
Analyzer as well as receiver mode
200 Hz、9 kHz、120 kHz (–6 dB)、
1 MHz (pulse bandwidth)With R&S®ESR-B29 option (in analyzer and receiver mode)
Increase by 10 Hz to 100 kHz
(–6 dB) in decimal stepspreconcentration
Can be turned off in analyzer mode
16 fixed filters
preamplifier
Can be turned on/off
1 kHz to 26.5 GHz
Measure time
Analyzer mode (scan time)
Frequency span = 0 Hz: 1 μs to 16 000 s
Frequency span ≥ 10 Hz (scan): 1 ms to 16 000 s
Frequency span ≥ 10 Hz (FFT): 7 μs to 16 000 sReceiver mode (step frequency sweep)
Receiver Mode (Time Domain Scan)50 μs to 100 s (depending on frequency)
50 μs to 100 s (depending on frequency subrange)Frequency step
Receiver mode (step frequency sweep)
Receiver Mode (Time Domain Scan)Minimum 1 Hz
0.25 × IF bandwidthdetector
Receiver mode
highest peak, lowest peak, quasi-peak, rms, average,
An average with metering time constants
(CISPR average), RMS average (CISPR-RMS)Display average noise level (DANL)
Receiver mode, nominal, average detector (AV), RF attenuation 0 dB, termination 50 Ω
Turn off the preamplifier
30 MHz < f < 1 GHz,, 120 kHz
< 8 dBμV
1 GHz < f < 3.6 GHz, 1 MHz
< 20 dBμV
3.6 GHz < f < 26.5 GHz, 1 MHz
< 26 dBμV
Turn on the preamplifier
30 MHz < f < 1 GHz,, 120 kHz
< –3 dBμV
1 GHz < f < 3.6 GHz, 1 MHz
< 9 dBμV
3.6 GHz < f < 26.5 GHz, 1 MHz
< 13 dBμV
The number of scan (trace) points
Analyzer Mode (Standard)
101 to 32 001
Analyzer Mode (EMI)
101 to 200 001
Receiver mode
Maximum 4 000 000
Real-time analyzer (optional)
801
Overall measurement uncertainty
CW signal, level 0 dB to –70 dB (below reference level), S/N > 20 dB, auto-scan time, RF attenuation 10 dB, 20 dB, 30 dB, 40 dB, preamplifier on, frequency span/RBW 95 % confidence, +20°C to +30°C
9 kHz ≤ f < 3.6 GHz
0.47 dB
3.6 GHz ≤ f ≤ 7 GHz
0.59 dB
7 GHz ≤ f ≤ 13.6 GHz
1.01 dB
13.6 GHz ≤ f ≤ 26.5 GHz
1.34 dB